Technical data

2-86 Chapter 2
Performance Tests
Performance Test Records
Agilent Technologies Company
Model 8753ES Option 011
Report Number
Serial Number Date
10. Receiver Trace Noise
Note: If your analyzer does not have Option 006, write “N/A” in all entries above 3 GHz.
Ratio IF
BW
Phase/Magnitude Measured Value Specification
(rms)
Test Frequency: 3 GHz
A/R 10 Hz Magnitude ≤0.001 dΒ
A/R 10 Hz Phase ≤0.006°
A/R 3 kHz Phase ≤0.038°
A/R 3 kHz Magnitude ≤0.006 dΒ
B/R 3 kHz Magnitude ≤0.006 dΒ
B/R 3 kHz Phase ≤0.038°
B/R 10 Hz Phase ≤0.006°
B/R 10 Hz Magnitude ≤0.001 dΒ
Test Frequency: 6 GHz
A/R 10 Hz Magnitude ≤0.002 dΒ
A/R 10 Hz Phase ≤0.012°
A/R 3 kHz Phase ≤0.070°
A/R 3 kHz Magnitude ≤0.010 dΒ
B/R 3 kHz Magnitude ≤0.010 dΒ
B/R 3 kHz Phase ≤0.070°
B/R 10 Hz Phase ≤0.012°
B/R 10 Hz Magnitude ≤0.002 dΒ