Technical data
Chapter 2 2-87
Performance Tests
Performance Test Records
Agilent Technologies Company
Model 8753ES Option 011 Report Number
Serial Number Date
11. Receiver Input Impedance
Note: If your analyzer does not have Option 006, write “N/A” in all entries above 3 GHz.
Frequency Range B Return
Loss (A/R)
A Return
Loss (B/R)
R Return
Loss (A/B)
Specification
(dB)
Measurement
Uncertainty
(dB)
300 kHz–2 MHz ≥20 ±0.6
2 MHz–1.3 GHz ≥24 ±0.6
1.3 GHz–3 GHz ≥19 ±0.6
3 GHz–6 GHz
a
a. Option 006 only.
≥15 ±2.0