Technical data
Chapter 1 1-3
Service Equipment and Analyzer Options
Required Tools and Equipment
Table 1-1 Service Test Equipment
Required Equipment Critical Specifications Recommended Model or
HP/Agilent Part Number
Use
a
Spectrum Analyzer Freq. Accuracy ±7 Hz 8563E A, T
Frequency Counter Frequency: 300 kHz–3 GHz
(6 GHz for Option 006)
5350B/51B/52B P
Synthesized Sweeper Maximum spurious input: <−30 dB
Residual FM: <20 kHz
83620A P
Oscilloscope Bandwidth: 100 MHz
Accuracy: 10%
any T
Digital Voltmeter Resolution: 10 mV any T
Power Meter (GPIB) 436A/437B/438A,
E4418A/B or E4419A/B
A, P, T
Power Sensor Frequency: 300 kHz–3 GHz, 50Ω 8482A A, P, T
Power Sensor
(for Option 006)
Frequency: 3 GHz–6 GHz 8481A A, P, T
S-Parameter Test Set Frequency Range: 300 kHz–3 GHz
for Option 006: 3 GHz–6 GHz
85046A
85047A
P
Transmission/Reflection
Test Set
Frequency Range: 300 kHz–3 GHz
Directivity: 30 dB
85044A P
Tool Kit No substitute 08753-60023 T
Photometer Tektronix J16 T
Photometer Probe Tektronix J6503 T
Light Occluder Tektronix 016-0305-00 T
Printer HP ThinkJet, DeskJet,
LaserJet
P
Floppy Disk one 3.5-inch formatted 1.44 MB any A
Calibration Kit 7 mm, 50Ω No substitute 85031B P
Calibration Kit Type-N, 50Ω No substitute 85032B P
Step Attenuator No substitute 8496A Opt. 003, H18 P
Fixed Attenuator Type-N, 3 dB 8491A Opt. 003 P
Fixed Attenuator (2) Type-N, 20 dB 8491A Opt. 020 A,P, T
Fixed Attenuator Type-N, 10 dB 8491A Opt. 010 A,P
Fixed Attenuator Type-N, 30 dB 8491A Opt. 030 A,P
RF Cable Set 12-inch, phase-matched 11500B A
RF Cable Kit Type-N, 50Ω, phase-matched 11851B A, P
RF Cable 24-inch, APC-7, 50 Ω 8120-4779 P