Technical data

Chapter 3 3-11
Adjustments and Correction Constants
RF Output Power Correction Constants (Test 47)
RF Output Power Correction Constants (Test 47)
This procedure adjusts several correction constants that can improve the output power
level accuracy of the internal source. They are related to the power level, power slope,
power slope offset, and the ALC roll-off factors among others.
Analyzer warmup time: 30 minutes.
Required Equipment and Tools
Procedure
1. If you just completed “Sampler Magnitude and Phase Correction Constants (Test 53),”
start at step 8 in this procedure.
2. Press .
3. Press . The default power meter
address is 13. Refer to the power meter manual as required to observe or change its
GPIB address.
4. Press to toggle between the HP/Agilent 438A/437 and 436A
power meters. Choose the appropriate model number. (Use the 438A/437 selection if the
power meter is an E4419B or E4418B.)
NOTE
If you are using the 438A power meter, connect the 8482A power sensor to
channel A, and the 8481A power sensor to channel B.
Description HP/Agilent Part or Model Number
Power meter 436A/437B/438A or E4418B/4419B
Power sensor 8482A
Power splitter (2) 11667A Option 001
Attenuator 20 dB 8491A Option 020
GPIB cable 10833A
RF cable 24-inch 11500B
Antistatic wrist strap 9300-1367
Antistatic wrist strap cord 9300-0980
Static-control table mat and earth ground wire 9300-0797
Additional Equipment for Option 006 Analyzers
Power sensor 8481A
Preset Local
SYSTEM CONTROLLER
Local
SET ADDRESSES ADDRESS: P MTR/GPIB
POWER MTR:438A/437