Technical data
Chapter 3 3-13
Adjustments and Correction Constants
RF Output Power Correction Constants (Test 47)
Figure 3-2 Setup A for RF Output Correction Constants
10.Press .
11.Record the power meter reading in the first column of Table 3-2.
12.Repeat the previous two steps at 50 MHz, 1.5 GHz, and 3 GHz.
13.For Option 006 Instruments Only: Make a measurement at 6 GHz by disconnecting
the 8482A (sensor A) from the power splitter, and replacing it with the 8481A (sensor
B).
• If you are using the 438A power meter, the 8481A should be connected to the meter's
channel B input.
• If you are using the 437B power meter, zero and calibrate the 8481A sensor.
14.Reconfigure the equipment as shown in Figure 3-3.
Table 3-2 Power Meter Readings
Setup A Reading
(First Reading)
Setup B Reading
(Second Reading)
Power Loss of #2
(Enter in Analyzer)
300 kHz: _______dB minus _______dB equals _______dB
50 MHz: _______dB minus _______dB equals _______dB
1.5 GHz: _______dB minus _______dB equals _______dB
3 GHz:_______ dB minus _______dB equals _______dB
6 GHz
a
: _______dB
a. For Option 006 instruments only.
minus _______dB equals _______dB
Sweep Setup
CW FREQ
300 k/m