Technical data

Chapter 3 3-13
Adjustments and Correction Constants
RF Output Power Correction Constants (Test 47)
Figure 3-2 Setup A for RF Output Correction Constants
10.Press .
11.Record the power meter reading in the first column of Table 3-2.
12.Repeat the previous two steps at 50 MHz, 1.5 GHz, and 3 GHz.
13.For Option 006 Instruments Only: Make a measurement at 6 GHz by disconnecting
the 8482A (sensor A) from the power splitter, and replacing it with the 8481A (sensor
B).
If you are using the 438A power meter, the 8481A should be connected to the meter's
channel B input.
If you are using the 437B power meter, zero and calibrate the 8481A sensor.
14.Reconfigure the equipment as shown in Figure 3-3.
Table 3-2 Power Meter Readings
Setup A Reading
(First Reading)
Setup B Reading
(Second Reading)
Power Loss of #2
(Enter in Analyzer)
300 kHz: _______dB minus _______dB equals _______dB
50 MHz: _______dB minus _______dB equals _______dB
1.5 GHz: _______dB minus _______dB equals _______dB
3 GHz:_______ dB minus _______dB equals _______dB
6 GHz
a
: _______dB
a. For Option 006 instruments only.
minus _______dB equals _______dB
Sweep Setup
CW FREQ
300 k/m