Technical data
Chapter 3 3-15
Adjustments and Correction Constants
RF Output Power Correction Constants (Test 47)
Source Correction Routine
20.Press to turn power loss ON.
21.Press .
22.Connect the equipment as shown in Figure 3-4, using splitter #2 and the power sensor
requested by the prompt.
Figure 3-4 Setup C for RF Output Correction Constants
23.Press .
24.Observe the analyzer display for the results of the adjustment routine:
• If the analyzer shows SOURCE Cor DONE, press and you have completed this
procedure.
• If the analyzer fails this routine, refer to Chapter 7 , “Source Troubleshooting.”
DONE RETURN PWR LOSS ON
RETURN TESTS
47 x1
EXECUTE TEST YES
CONTINUE
Preset