Technical data

Chapter 3 3-17
Adjustments and Correction Constants
IF Amplifier Correction Constants (Test 51)
4. Press and observe the analyzer for the results of the adjustment routine:
•If DONE is displayed, you have completed this procedure.
•If FAIL is displayed, check that the RF cable is connected between Port 1 and Port 2.
Then, repeat this adjustment routine.
If the analyzer continues to fail the adjustment routine, refer to Chapter 6 , “Digital
Control Troubleshooting.”
CONTINUE