Technical data

3-20 Chapter 3
Adjustments and Correction Constants
Sampler Magnitude and Phase Correction Constants (Test 53)
Power Sensor Calibration Factor Entry
5. Press
to access the calibration factor menu for power sensor A.
6. Build a table of up to 55 points (55 frequencies with their calibration factors). To enter
each point, follow these steps:
a. Press .
b. Input a frequency value and then press the appropriate key ( , , or ).
c. Press and enter the calibration factor percentage that corresponds to
the frequency you entered.
The cal factor and frequency values are found on the back of the sensor. If you make
a mistake, press and re-enter the correct value.
d. Press to complete the data entry for each point.
NOTE
The following terms are part of the sensor calibration menu.
7. For Option 006 Instruments Only: Press to create a
power sensor calibration table for power sensor B (8481A), using the softkeys
mentioned previously. Since sensor B is only used for 3 GHz to 6 GHz measurements,
you only need to input calibration factors for this frequency range.
8. Zero and calibrate the power meter and power sensor.
Update Sampler Correction Constants
9. Preset, zero, and calibrate the power meter and 8482A (sensor A).
10.Press .
11.Connect the equipment as shown in Figure 3-6.
allows you to select a frequency point.
allows you to edit or change a previously entered value.
allows you to delete a point from the sensor cal factor table.
allows you to add a point into the sensor cal factor table.
allows you to erase the entire sensor cal factor table.
allows you to complete the points entry of the sensor cal
factor table.
System
SERVICE MENU TEST OPTIONS LOSS/SENSR LISTS
CAL FACTOR SENSOR A
ADD FREQUENCY
G/n M/ยต k/m
CAL FACTOR
DONE
SEGMENT
EDIT
DELETE
ADD
CLEAR LIST
DONE
CAL FACTOR SENSOR B
System
SERVICE MENU TESTS
53 x1
EXECUTE TEST YES