Technical data

3-24 Chapter 3
Adjustments and Correction Constants
Sampler Magnitude and Phase Correction Constants (Test 53)
9. Update the offsets for each sampler:
A-Channel Sampler
a. Access the first address by pressing
.
b. Enter the new value for the A-sampler at the accessed address by pressing
[new value for A] .
c. Press for the analyzer to use the new values.
B-Channel Sampler
a. Access the first address by pressing
.
b. Enter the new value for the B-sampler at the accessed address by pressing
[new value for B] .
c. Press for the analyzer to use the new values.
R-Channel Sampler
a. Access the first address by pressing
.
b. Enter the new value for the R-sampler at the accessed address by pressing
[new value for R] .
c. Press for the analyzer to use the new values.
10.Repeat this adjustment procedure.
11.If the analyzer continues to fail this adjustment routine, refer to Chapter 8 , “Receiver
Troubleshooting.”