Technical data

4-18 Chapter 4
Start Troubleshooting Here
Receiver
Figure 4-9 Typical Measurement Trace
If the source is working, but the R, A, or B input traces appear to be in error, refer to
Chapter 8 , “Receiver Troubleshooting.”
The following symptoms may also indicate receiver failure.
Receiver Error Messages
CAUTION: OVERLOAD ON INPUT A; POWER REDUCED
CAUTION: OVERLOAD ON INPUT B; POWER REDUCED
CAUTION: OVERLOAD ON INPUT R; POWER REDUCED
The error messages above indicate that you have exceeded approximately +14 dBm at one
of the test ports. The RF output power is automatically reduced to 85 dBm. The
annotation P appears in the left margin of the display to indicate that the power trip
function has been activated. When this occurs, press and enter a lower power
level. Press to switch on the power again.
Faulty Data
Any trace data that appears to be below the noise floor of the analyzer (100 dBm) is
indicative of a receiver failure.
Power
SOURCE PWR ON