Technical data

5-6 Chapter 5
Power Supply Troubleshooting
Start Here
Measure the Post Regulator Voltages
Measure the DC voltages on the test points of A8 with a voltmeter. Refer to Figure 5-3 for
test point locations and Table 5-3 for supply voltages and limits.
Figure 5-3 A8 Post Regulator Test Point Locations
Table 5-1. A8 Post Regulator Test Point Voltages
TP Supply Range
1 +65 V (not used) +64.6 to +65.4
2AGND n/a
3 +5 VD +4.9 to +5.3
4SDIS n/a
5 15 V 14.4 to 15.6
6 12.6 VPP (probe power) 12.1 to 12.91
7 +15 V +14.5 to +15.5
8 +5 VU +5.05 to +5.35
9 5.2 V 5.0 to 5.4
10 +22 V +21.3 to +22.7
11 +6 V +5.8 to +6.2