Technical data

2-4 Chapter 2
Performance Tests
1. Source Frequency Range and Accuracy
1. Source Frequency Range and Accuracy
Perform this test to verify the frequency accuracy of the analyzer over its entire operating
frequency range. A frequency counter is used to determine the analyzer’s output frequency.
Analyzer warm-up time: 30 minutes
Specifications
Required Equipment
Procedure
1. Connect the equipment as shown in Figure 2-1.
Frequency Range
Frequency Accuracy
a
a. At 25 °C ±5 °C.
300 kHz to 3 GHz ±10 ppm
30 kHz to 6 GHz
b
b. Only for analyzers with Option 006.
±10 ppm
Description HP/Agilent Part or Model Number
Frequency Counter 5350B/51B/52B
RF cable set, 50 type-N 11851B
2-way power splitter, 50 11667A Option 001
Adapter, type-N (f) to BNC (m) 1250-0077