2. Source Power Range, Linearity, and Accuracy
Figure 2-3 Path Loss Calibration Test Setup (#2)
8. Press . Set the power meter calibration factor for
this CW frequency.
9. Write the power meter reading in the “Second Value” column on the performance test
record.
10.Calculate the path loss through the power splitter with this formula:
First Value − Second Value = Path Loss
11.Write the result on the performance test record.
12.Repeat steps 8 through 11 for the other CW frequencies listed on the performance test
record.