Technical data
Chapter 2 2-11
Performance Tests
2. Source Power Range, Linearity, and Accuracy
In Case of Difficulty
1. Ensure that the power meter and power sensor are operating to specification.
2. Inspect the power splitter connectors. Poor match at these connections can generate
power reflections that can cause the analyzer to appear to be out of specification.
3. Inspect the analyzer RF OUT connector for damage.
4. If any test fails, perform “RF Output Power Correction Constants (Test 47)” on
page 3-11.
5. For a source failure, refer to Chapter 7 , “Source Troubleshooting,” for troubleshooting
information.