Technical data

2-12 Chapter 2
Performance Tests
3. Receiver Minimum R Channel Level
3. Receiver Minimum R Channel Level
Perform this test to verify the minimum R channel input power level at which phase lock
can be accomplished.
Analyzer warmup time: 30 minutes.
Specifications
Required Equipment
Procedure
1. Connect the equipment as shown in Figure 2-5.
Figure 2-5 Receiver Minimum R Channel Level Test Setup
Frequency Range Minimum R Channel Level
300 kHz–3 GHz 35 dBm
3 GHz–6 GHz
a
a. Only for analyzers with Option 006.
30 dBm
Description HP/Agilent Part or Model Number
Cable Type-N (m) to Type-N (m) 24-inch 11500B
Attenuator, 30 dB 8491A Option 030