Technical data

9-2 Chapter 9
Accessories Troubleshooting
Accessories Troubleshooting
Accessories Troubleshooting
Use this procedure only if you have read Chapter 4 , “Start Troubleshooting Here. Follow
the procedures in the order given, unless instructed otherwise.
Measurement failures can be divided into two categories:
Failures which don't affect the normal functioning of the analyzer but render incorrect
measurement data.
Failures which impede the normal functioning of the analyzer or prohibit the use of a
feature.
This chapter addresses the first category of failures which are usually caused by the
following:
operator errors
faulty calibration devices or connectors
bad cables or adapters
improper calibration techniques
RF cabling problems within the test set.
These failures are checked using the following procedures:
“Inspect the Accessories” on page 9-4
“Inspect the Error Terms” on page 9-5
“Test Set Troubleshooting” on page 9-8