Technical data

Chapter 10 10-35
Service Key Menus and Error Messages
Service Key Menus
Node 20 IF Det 1 (IF after 30 MHz filter)
Perform step A11 and then press
.
The trace should be a flat line across the entire frequency band at least 0.5 V greater than
Vbb (node 14). The correct trace indicates the presence of IF after the first 30 MHz
filter/limiter.
Figure 10-14 Analog Bus Node 20
A12 Reference To observe the A12 analog bus nodes perform step A12, below. Then
follow the node-specific instructions.
Step A12: Press
.
Node 21 100 kHz (100 kHz reference frequency)
Perform step A12 and then press
. This node counts the A12 100 kHz reference signal that is
used on A13 (the fractional-N analog assembly) as a reference frequency for the phase
detector.
Node 22 A12 Gnd 1 (ground)
Meas
ANALOG IN
20 x1 Scale Ref 0.3 x1
REFERENCE VALUE
1.29 x1
Preset Meas
ANALOG IN
Marker System
SERVICE MENU
ANALOG BUS ON
Format
MORE REAL
Meas
ANALOG IN
21 x1
COUNTER: ANALOG BUS