Technical data
Chapter 2 2-17
Performance Tests
5. Receiver Channel Noise Floor Level (Only for Analyzers without Option 006)
Figure 2-7 Receiver Channel Noise Floor Level Test Setup
2. Press .
Receiver Channel A Noise Floor Level with 3 kHz IF BW
3. Press .
4. Press
.
5. When the analyzer finishes the sweep, notice the mean value, which appears on the
analyzer display.
6. Use the following equation to convert the linear magnitude mean value to log
magnitude:
Power (dBm) = 20 × [log
10
(linear magnitude mean value)]
7. Write this calculated value in the performance test record in dBm.
Receiver Channel A Noise Floor Level with 10 Hz IF BW
8. Press to change the IF bandwidth to 10 Hz.
9. Press .
10.When the analyzer finishes the sweep, notice the mean value, which appears on the
analyzer display.
11.Use the following equation to convert the linear magnitude mean value to log
magnitude:
Power (dBm) = 20 × [log
10
(linear magnitude mean value)]
12.Write this calculated value in the performance test record in dBm.
Preset Power −10 x1
Meas
A
Format
LIN MAG
Scale Ref
AUTO SCALE
Marker Fctn
MARKER MODE MENU MKR STATS ON
Sweep Setup
TRIGGER MENU SINGLE
Av
g
IF BW
10 x1
Sweep Setup
TRIGGER MENU SINGLE