Technical data
11-4 Chapter 11
Error Terms
Full Two-Port Error-Correction Procedure
Full Two-Port Error-Correction Procedure
NOTE
This is the most accurate error-correction procedure for the
analyzer. Since the analyzer takes both forward and reverse sweeps, this
procedure takes more time than the other correction procedures.
1. Set any measurement parameters that you want for the device measurement: power,
format, number of points, IF bandwidth.
2. To access the measurement correction menus, press .
3. Press .
4. To select the correction type, press:
5. Connect a shielded open circuit to PORT 1.
Figure 11-1 Standard Connections for Full Two-Port Error-Correction
6. To measure the standard, when the displayed trace has settled, press
.
The analyzer underlines the softkey after it measures the standard.
7. Disconnect the open, and connect a short circuit to PORT 1.
8. To measure the device, when the displayed trace has settled, press
.
The analyzer underlines the softkey after it measures the standard.
Cal
CAL KIT SELECT CAL KIT 7 mm 85031 RETURN
CALIBRATE MENU FULL 2-PORT REFLECTION
FORWARD: OPEN
OPEN
FORWARD: SHORT
SHORT