Technical data
11-6 Chapter 11
Error Terms
Full Two-Port Error-Correction Procedure
c. Press
.
d. Return the averaging to the original state of the measurement, and press
.
17.To compute the error coefficients, press .
The analyzer displays the corrected measurement trace. The analyzer also shows the
notation Cor at the left of the screen, indicating that error-correction is on.
NOTE
You can save or store the measurement correction to use for later
measurements. Use the menus under , or refer to your analyzer’s
user’s guide for procedures.
18.This completes the full two-port correction procedure. You can connect and measure
your device under test.
Cal
RESUME CAL SEQUENCE FWD ISOL'N ISOL'N STD
REV ISOL'N ISOL'N STD ISOLATION DONE
Cal
RESUME CAL SEQUENCE
DONE 2-PORT CAL
Save/Recall