Technical data

Chapter 11 11-7
Error Terms
Full Two-Port Error-Correction Procedure
Table 11-1 Calibration Coefficient Terms and Tests
Calibration
Coefficient
Calibration Type Test Number
Response Response and
Isolation
a
a.
Response and Isolation cal yields: E
X
or E
T
if a transmission parameter (S
21
, S
12
) or E
D
or
E
R
if a reflection parameter (S
11
, S
22
).
1-port
2-port
b
b.
One-path, 2-port cal duplicates arrays 1 to 6 in arrays 7 to 12.
1E
R
or E
T
E
X
(E
D
)E
D
E
DF
32
2E
T
(E
R
)E
S
E
SF
33
3E
R
E
RF
34
4E
XF
35
5E
LF
36
6E
TF
37
7E
DR
38
8E
SR
39
9E
RR
40
10 E
XR
41
11 E
LR
42
12 E
TR
43
Meaning of first subscript: Meaning of second subscript:
D: directivity F: forward
S: source match R: reverse
R: reflection tracking
X: crosstalk or isolation
L: load match
T: transmission tracking