Technical data
11-8 Chapter 11
Error Terms
Full Two-Port Error-Correction Procedure
Error Term Inspection
NOTE
If the correction is not active, press .
1. Press .
The analyzer copies the first calibration measurement trace for the selected error term
into memory and then displays it. Table 11-1 lists the test numbers.
2. Press and adjust the scale and reference to study the error term trace.
3. Press and use the marker functions to determine the error term
magnitude.
4. Compare the displayed measurement trace to the traces shown in “Error Term
Descriptions” on page 11-9, and to previously measured data.
5. Make a hardcopy of the measurement results:
a. Connect a printing or plotting peripheral to the analyzer.
b. Press and select the
appropriate peripheral to verify that the GPIB address is set correctly on the
analyzer.
c. Press and then choose either or .
d. Press and title each data trace so that you can identify it
later.
For detailed information on creating hardcopies, refer to your analyzer’s user’s guide.