Technical data

Chapter 2 2-19
Performance Tests
6. Receiver Channel Noise Floor Level (Only for Analyzers with Option 006)
6. Receiver Channel Noise Floor Level
(Only for Analyzers with Option 006)
Perform this test to determine the analyzer's receiver channel noise floor levels.
Analyzer warmup time: 30 minutes.
Specifications
Required Equipment
Procedure
1. Connect the equipment as shown in Figure 2-8.
Frequency Range Receiver Input IF Bandwidth Average Noise Level
300 kHz–3.0 GHz A 3 kHz 90 dBm
300 kHz–3.0 GHz A 10 Hz 110 dBm
300 kHz–3.0 GHz B 3 kHz 90 dBm
300 kHz–3.0 GHz B 10 Hz 110 dBm
3.0 GHz–6.0 GHz A 3 kHz 85 dBm
3.0 GHz–6.0 GHz A 10 Hz 105 dBm
3.0 GHz–6.0 GHz B 3 kHz 85 dBm
3.0 GHz–6.0 GHz B 10 Hz 105 dBm
Description HP/Agilent Part or Model Number
Attenuator, 20 dB 8491A Option 020
Termination (2), 50
909F Option 012
a
a. Or use the terminations included in the 85032B calibration kit.
Cable, 50 type-N, 24-inch Part of the 11851B Cable Set