Technical data

Chapter 2 2-23
Performance Tests
7. Receiver Magnitude Frequency Response
7. Receiver Magnitude Frequency Response
Perform this test to verify how well the analyzer transfers information from the RF to IF,
and how accurately it processes and displays that information.
Analyzer warmup time: 30 minutes.
Specifications
Required Equipment
Procedure
Input R Magnitude Frequency Response
1. Connect the equipment as shown in Figure 2-9.
Frequency Range
Magnitude Frequency Response
a
a. At 25 °C ± 5 °C; A, B, R, 10 dBm input.
300 kHz–3 GHz ± 1 dB
3 GHz–6 GHz
b
b. Only for analyzers with Option 006.
± 2 dB
Description HP/Agilent Part or Model Number
Power Meter 436A/437B/438A or E4418B/4419B
Power Sensor 8482A
2-Way power splitter, 50 11667A Option 001
Attenuator, 10 dB 8491A Option 010
RF cable set, 50 type-N 11851B
Adapter, type-N (m) to type-N (m) 1250-1475
Adapter, type-N (f) to type-N (f) 1250-1472
Additional Equipment for Option 006 Analyzers
Power Sensor 8481A