Technical data

Chapter 2 2-25
Performance Tests
7. Receiver Magnitude Frequency Response
Figure 2-10 Magnitude Frequency Response Test Setup (Receiver Input A)
9. Press to measure the power at the receiver input A channel.
10.Press .
11.Write the marker reading in the A Input Power” column on the performance test
record.
12.Repeat steps 10 and 11 for the other CW frequencies listed on the performance test
record.
Input B Magnitude Frequency Response
13.Reconfigure the equipment as shown in Figure 2-11.
Figure 2-11 Magnitude Frequency Response Test Setup (Receiver Input B)
Meas
A
Sweep Setup
CW FREQ
300 k/m