Technical data
Chapter 2 2-25
Performance Tests
7. Receiver Magnitude Frequency Response
Figure 2-10 Magnitude Frequency Response Test Setup (Receiver Input A)
9. Press to measure the power at the receiver input A channel.
10.Press .
11.Write the marker reading in the “A Input Power” column on the performance test
record.
12.Repeat steps 10 and 11 for the other CW frequencies listed on the performance test
record.
Input B Magnitude Frequency Response
13.Reconfigure the equipment as shown in Figure 2-11.
Figure 2-11 Magnitude Frequency Response Test Setup (Receiver Input B)
Meas
A
Sweep Setup
CW FREQ
300 k/m