Technical data
2-28 Chapter 2
Performance Tests
8. Receiver Phase Frequency Response (Ratio)
Figure 2-12 Phase Frequency Response Test Setup
From 300 kHz to 3 GHz
2. Press .
3. Press .
4. Press .
5. Only For Analyzers with Option 006: Press .
6. Press .
7. Press . Turn the analyzer front panel
knob to vary the electrical delay until the trace is in the most linear horizontal position.
8. Press . Press
. Press
. Record the marker 1 or marker 2 value—whichever has the greatest
absolute value—in the “Measured Value” column of the performance test record.
From 3 GHz to 6 GHz (Option 006)
9. Press .
10.Press . Turn the analyzer front panel knob
to vary the electrical delay until the trace is in the most linear horizontal position.
11.Press . Press
. Press
. Record the marker 1 or marker 2 value—whichever has the greatest
absolute value—in the “Measured Value” column of the performance test record.