Performance Tests
8. Receiver Phase Frequency Response (Ratio)
B/R Phase Frequency Response
12.Connect the second type-N cable from the RF cable set to port B.
From 300 kHz to 3 GHz
13.Press .
14.Press .
15.Only For Analyzers with Option 006: Press .
16.Press . Turn the analyzer front panel
knob to vary the electrical delay until the trace is in the most linear horizontal position.
17.Press . Press
. Press
. Record the marker 1 or marker 2 value—whichever has the greatest
absolute value—in the “Measured Value” column of the performance test record.
From 3 GHz to 6 GHz (Option 006)
18.Press .
19.Press . Turn the analyzer front panel knob
to vary the electrical delay until the trace is in the most linear horizontal position.
20.Press . Press
. Press
. Record the marker 1 or marker 2
value—whichever has the greatest absolute value—in the “Measured Value” column of
the performance test record.
A/B Phase Frequency Response
21.Connect the equipment as shown in Figure 2-13.
CONTINUOUS
Start 300 k/m
Meas
B/R
Format
PHASE
Stop 3 G/n
Scale Ref 0.6 x1
ELECTRICAL DELAY
Sweep Setup
TRIGGER MENU SINGLE
Marker
MARKER 1
Marker Search
SEARCH: MAX
Marker
MARKER 2
Marker Search
SEARCH: MIN
CONTINUOUS
Start 3 G/n Stop 6 G/n
Scale Ref 5 x1
ELECTRICAL DELAY
Sweep Setup
TRIGGER MENU SINGLE
Marker
MARKER 1
Marker Search
SEARCH: MAX
Marker
MARKER 2
Marker Search
SEARCH: MIN