Technical data
2-30 Chapter 2
Performance Tests
8. Receiver Phase Frequency Response (Ratio)
Figure 2-13 Phase Frequency Response Test Setup
From 300 kHz to 3 GHz
22.Press .
23.Press .
24.Press .
25.Only For Analyzers with Option 006: Press .
26.Press . Turn the analyzer front panel
knob to vary the electrical delay until the trace is in the most linear horizontal position.
27.Press . Press
. Press
. Record the marker 1 or marker 2 value—whichever has the greatest
absolute value—in the “Measured Value” column of the performance test record.
From 3 GHz to 6 GHz (Option 006)
28.Press .
29.Press . Turn the analyzer front panel knob
to vary the electrical delay until the trace is in the most linear horizontal position.
30.Press . Press
. Press
. Record the marker 1 or marker 2 value—whichever has the greatest
absolute value—in the “Measured Value” column of the performance test record.
CONTINUOUS
Start 300 k/m
Avg
IF BW
300 x1
Meas
A/B
Format
PHASE
Stop 3 G/n
Scale Ref 0.6 x1
ELECTRICAL DELAY
Sweep Setup
TRIGGER MENU SINGLE
Marker
MARKER 1
Marker Search
SEARCH: MAX
Marker
MARKER 2
Marker Search
SEARCH: MIN
CONTINUOUS
Start 3 G/n Stop 6 G/n
Scale Ref 5 x1
ELECTRICAL DELAY
Sweep Setup
TRIGGER MENU SINGLE
Marker
MARKER 1
Marker Search
SEARCH: MAX
Marker
MARKER 2
Marker Search
SEARCH: MIN