Technical data

Chapter 2 2-33
Performance Tests
9. Receiver Input Crosstalk
Figure 2-14 R into A and R into B Crosstalk Test Setup
2. Press
.
3. Press to get a better scaling of the data trace.
4. Press .
5. Press
.
6. When the analyzer finishes the number of sweeps, write the marker value, which
appears on the analyzer display, on the performance test record.
R into A Crosstalk from 1 GHz to 3 GHz
7. Press .
8. When the analyzer finishes the number of sweeps, write the marker value, which
appears on the analyzer display, on the performance test record.
(Only for Option 006 Analyzers) R into A Crosstalk from 3 GHz to 4.5 GHz
9. Press .
10.When the analyzer finishes the number of sweeps, write the marker value, which
appears on the analyzer display, on the performance test record.
(Only for Option 006 Analyzers) R into A Crosstalk from 4.5 GHz to 6 GHz
11.Press .
12.When the analyzer finishes the number of sweeps, write the marker value, which
appears on the analyzer display, on the performance test record.
R into B Crosstalk from 300 kHz to 1 GHz
13.Press .
Preset Power 6 x1 Avg
AVERAGING FACTOR
5 x1
AVERAGING ON IF BW
10 x1
Scale Ref 25 x1
Marker Search
TRACKING ON SEARCH:MAX
Stop 1 G/n Sweep Setup
TRIGGER MENU NUMBER of GROUPS
5
x1
Start 1 G/n Stop 3 G/n
NUMBER of GROUPS
5 x1
Start 3 G/n Stop 4.5 G/n
NUMBER of GROUPS
5 x1
Start 4.5 G/n Stop 6 G/n
NUMBER of GROUPS
5 x1
Start 300 k/m Stop 1 G/n