Technical data

2-34 Chapter 2
Performance Tests
9. Receiver Input Crosstalk
14.Press
.
15.When the analyzer finishes the number of sweeps, write the marker value, which
appears on the analyzer display, on the performance test record.
R into B Crosstalk from 1 GHz to 3 GHz
16.Press
.
17.When the analyzer finishes the number of sweeps, write the marker value, which
appears on the analyzer display, on the performance test record.
(Only for Option 006 Analyzers) R into B Crosstalk from 3 GHz to 4.5 GHz
18.Press
.
19.When the analyzer finishes the number of sweeps, write the marker value, which
appears on the analyzer display, on the performance test record.
(Only for Option 006 Analyzers) R into B Crosstalk from 4.5 GHz to 6 GHz
20.Press
.
21.When the analyzer finishes the number of sweeps, write the marker value, which
appears on the analyzer display, on the performance test record.
B into A Crosstalk from 300 kHz to 1 GHz
22.Connect the equipment as shown in Figure 2-15.
Figure 2-15 B into A Receiver Input Crosstalk Test Setup
23.Press .
Meas
B/R
Sweep Setup
TRIGGER MENU NUMBER of GROUPS
5
x1
Start 1 G/n Stop 3 G/n
TRIGGER MENU NUMBER of GROUPS
5 x1
Start 3 G/n Stop 4.5 G/n
TRIGGER MENU NUMBER of GROUPS
5 x1
Start 4.5 G/n Stop 6 G/n
TRIGGER MENU NUMBER of GROUPS
5 x1
Start 300 k/m Stop 1 G/n