Technical data
Chapter 2 2-35
Performance Tests
9. Receiver Input Crosstalk
24.Press
.
25.When the analyzer finishes the number of sweeps, write the marker value, which
appears on the analyzer display, on the performance test record.
B into A Crosstalk from 1 GHz to 3 GHz
26.Press
.
27.When the analyzer finishes the number of sweeps, write the marker value, which
appears on the analyzer display, on the performance test record.
(Only for Option 006 Analyzers) B into A Crosstalk from 3 GHz to 4.5 GHz
28.Press
.
29.When the analyzer finishes the number of sweeps, write the marker value, which
appears on the analyzer display, on the performance test record.
(Only for Option 006 Analyzers) B into A Crosstalk from
4.5 GHz to 6 GHz
30.Press
.
31.When the analyzer finishes the number of sweeps, write the marker value, which
appears on the analyzer display, on the performance test record.
A into B Crosstalk from 300 kHz to 1 GHz
32.Reconfigure the equipment as shown in Figure 2-16.
Figure 2-16 A into B Receiver Input Crosstalk Test Setup
Meas
A/B
Sweep Setup
TRIGGER MENU NUMBER of GROUPS
5
x1
Start 1 G/n Stop 3 G/n
TRIGGER MENU NUMBER of GROUPS
5 x1
Start 3 G/n Stop 4.5 G/n
TRIGGER MENU NUMBER of GROUPS
5 x1
Start 4.5 G/n Stop 6 G/n
TRIGGER MENU NUMBER of GROUPS
5 x1