Technical data
Chapter 2 2-39
Performance Tests
10. Receiver Trace Noise
Figure 2-17 Receiver Trace Noise Test Setup
2. Press .
3. Press .
4. Press to enable the analyzer
marker statistics feature.
A/R Trace Noise Magnitude IF BW = 10 Hz
5. Press .
6. Press .
7. Press . Wait for the sweep to finish as
indicated by the Hld notation on the left side of the display.
8. Press .
9. Write the s. dev (standard deviation) value shown, which appears on the analyzer
display, on the performance test record.
A/R Trace Noise Phase IF BW = 10 Hz
10.Press .
11.Press . Wait for the sweep to finish as
indicated by the Hld notation on the left side of the display.
12.Press .
13.Write the s. dev (standard deviation) value shown, which appears on the analyzer
display, on the performance test record.
Preset Power 10 x1 Sweep Setup
NUMBER of POINTS
51 x1
CW FREQ
3 G/n
Marker Fctn
MARKER MODE MENU MKR STATS ON
Avg
IF BW
10 x1
Meas
A/R
Sweep Setup
TRIGGER MENU SINGLE
Scale Ref
AUTO SCALE
Format
PHASE
Sweep Setup
TRIGGER MENU SINGLE
Scale Ref
AUTO SCALE