10. Receiver Trace Noise
A/R Trace Noise Phase IF BW = 3 kHz
14.Press .
15.Press . Wait for the sweep to finish as
indicated by the Hld notation on the left side of the display.
16.Press .
17.Write the s. dev (standard deviation) value shown, which appears on the analyzer
display, on the performance test record.
A/R Trace Noise Magnitude IF BW = 3 kHz
18.Press .
19.Press . Wait for the sweep to finish as
indicated by the Hld notation on the left side of the display.
20.Press .
21.Write the s. dev (standard deviation) value shown, which appears on the analyzer
display, on the performance test record.
B/R Trace Noise Magnitude IF BW = 3 kHz
22.Disconnect the RF cable from input A, and connect it to input B.
23.Press .
24.Press . Wait for the sweep to finish as
indicated by the Hld notation on the left side of the display.
25.Press .
26.Write the s. dev (standard deviation) value shown, which appears on the analyzer
display, on the performance test record.
B/R Trace Noise Phase IF BW = 3 kHz
27.Press .
28.Press . Wait for the sweep to finish as
indicated by the Hld notation on the left side of the display.
29.Press .
30.Write the s. dev (standard deviation) value shown, which appears on the analyzer
display, on the performance test record.
Avg
IF BW
3 k/m
Sweep Setup
TRIGGER MENU SINGLE
Scale Ref
AUTO SCALE
Format
LOG MAG
Sweep Setup
TRIGGER MENU SINGLE
Scale Ref
AUTO SCALE
Meas
B/R
Sweep Setup
TRIGGER MENU SINGLE
Scale Ref
AUTO SCALE
Format
PHASE
Sweep Setup
TRIGGER MENU SINGLE
Scale Ref