Technical data
Chapter 2 2-49
Performance Tests
12. Receiver Magnitude Dynamic Accuracy
Required Equipment
NOTE
The HP/Agilent 8496A step attenuator (Option H19) comes with a special
calibration that supports the measurement uncertainties expressed in the
test record for this performance test.
The special calibration consists of two measurements. The first is a
measurement of the attenuation at each step. The data reported for this
measurement have the following uncertainties:
• ±0.006 dB from 0 to 40 dB
• ±0.015 dB from >40 to 80 dB
• ±0.025 dB from >80 to 90 dB
• ±0.05 dB >90 dB
The second calibration measurement characterizes match stability between
attenuator settings for each attenuator port. The vector difference of S
11
or
(S
22
) between the reference attenuation step and all the other steps is
measured. The magnitude of this difference is certified to be <0.0316
(>30 dB).
Description HP/Agilent Part or Model Number
Power Meter 436A/437B/438A or E4418B/E4419B
Power Sensor 8482A
Step Attenuator, 110 dB 8496A Option H19 (See note below.)
Cable (2), APC-7, 50 Ω, 24-inch 8120-4779
Cable, GPIB 10833A
Calibration Kit, 7-mm 85031B
S-Parameter Test Set 85047A/85406A
Additional Equipment for
the 85046B Test Set (75 Ω)
Minimum Loss Pad (2), 50 Ω to 75 Ω 11852B
Adapter (2), APC-7 to Type-N (m) 11525A