Technical data
2-52 Chapter 2
Performance Tests
12. Receiver Magnitude Dynamic Accuracy
Figure 2-22 Power Meter Calibration
8. Set the 8496A to 20 dB.
9. Set the following analyzer parameters:
•
•
•
•
10.Set up the analyzer for power meter calibration:
a. Select the analyzer as the system controller:
•
•
b. Set the power meter's address:
•
•
c. Select the appropriate power meter by pressing until the correct
model number is displayed. (Use the 438A/437 selection if the power meter is an
HP/Agilent E4419B or E4418B.)
d. Select the calibration kit and enter the power sensor calibration data.
•
• (enter
the power sensor calibration data for 30 MHz)
Preset Sweep Setup
CW FREQ
30 M/µ
NUMBER of POINTS
51 x1
Power 10 x1
Avg
IF BW
10 x1
Local
SYSTEM CONTROLLER
SET ADDRESSES
ADDRESS: P MTR/GPIB
13 x1
POWER MTR
Cal
CAL KIT SELECT CAL KIT 7mm 85031
Cal
PWRMTR CAL LOSS/SENSOR LISTS CAL FACTOR SENSOR A
DONE