Technical data

2-52 Chapter 2
Performance Tests
12. Receiver Magnitude Dynamic Accuracy
Figure 2-22 Power Meter Calibration
8. Set the 8496A to 20 dB.
9. Set the following analyzer parameters:
10.Set up the analyzer for power meter calibration:
a. Select the analyzer as the system controller:
b. Set the power meter's address:
c. Select the appropriate power meter by pressing until the correct
model number is displayed. (Use the 438A/437 selection if the power meter is an
HP/Agilent E4419B or E4418B.)
d. Select the calibration kit and enter the power sensor calibration data.
(enter
the power sensor calibration data for 30 MHz)
Preset Sweep Setup
CW FREQ
30 M/µ
NUMBER of POINTS
51 x1
Power 10 x1
Avg
IF BW
10 x1
Local
SYSTEM CONTROLLER
SET ADDRESSES
ADDRESS: P MTR/GPIB
13 x1
POWER MTR
Cal
CAL KIT SELECT CAL KIT 7mm 85031
Cal
PWRMTR CAL LOSS/SENSOR LISTS CAL FACTOR SENSOR A
DONE