Technical data

2-56 Chapter 2
Performance Tests
12. Receiver Magnitude Dynamic Accuracy
4. Take a power meter calibration sweep.
5. Verify that the power meter reads approximately 30 dBm.
Full 2-Port Calibration
6. Connect the equipment as shown in the Figure 2-26 Be sure to reconnect the cable to
the A channel connectors.
Figure 2-26 Full 2-Port Calibration
7. Perform a full 2-port error correction with isolation.
8. Save the results of the new cal set.
Measure Channel A Magnitude Dynamic Accuracy
9. Set the step attenuator to 20 dB.
10.To set up the dynamic accuracy measurement, press the following:
11.Wait for the sweep to finish, then press .
12.Set the step attenuator to 0 dB.
13.Press .
14.Press . Write the mean value (which appears on the
analyzer’s display) in the column marked “Real Part” in the performance test record.
Cal
PWRMTR CAL
30 x1
ONE SWEEP TAKE CAL SWEEP
Display
DATA
Sweep Setup
TRIGGER MENU SINGLE
Display
DATA MEMORY DATA/MEM
Sweep Setup
TRIGGER MENU SINGLE
Format
MORE REAL