Technical data

2-64 Chapter 2
Performance Tests
14. Receiver Phase Compression
12.Press . (This step is only necessary for the first frequency.)
13.Press .
14.Press .
15. Determine which value of marker 2 and marker 3 has the largest absolute value.
Record this value in the column of the test record labeled “Measured Value.”
16.Press . Set the calibration factor on the power
meter for this CW frequency (or the next CW frequency from the performance test
record).
17.Press
.
18.Press . Use the analyzer's front panel knob to adjust the source power for a
power meter reading of 20.00 dBm. Write the source power value, which appears on
the analyzer's display, in the “Start Power” column of the performance test record.
19.Press . Use the analyzer's front panel knob to adjust the source power for a
power meter reading of 0.00 dBm. Write the source power value in the “Stop Power
column of the performance test record.
20.Press . Use the
analyzer's keypad to enter the “Start Power” value as determined in step 18.
21.Press and enter the “Stop Power” value as determined in step 19.
22.Press . At the end of the sweep, press
.
23.Repeat steps 11 to 22 (omitting step 12) for the other CW frequencies listed in the
performance test record.