Technical data
2-66 Chapter 2
Performance Tests
15. Source and Receiver Harmonics (Option 002 Only)
15. Source and Receiver Harmonics (Option 002 Only)
Perform this test to determine the 2nd and 3rd harmonics of the source and receiver.
Analyzer warmup time: 30 minutes
Specifications
Required Equipment
Source Harmonics
a
a. At +20 dBm output level (or +18 dBm for Option 006).
Power Level
2nd Harmonic 16 MHz–3 GHz < −25 dBc
3rd Harmonic 16 MHz–2 GHz <−25 dBc
Receiver Harmonics
b
b. At 0 dBm input level.
2nd Harmonic < −15 dBc
3rd Harmonic <−30 dBc
Source and Receiver Harmonics
2nd Harmonic <−15 dBc
3rd Harmonic <−30 dBc
Description HP/Agilent Model Number
2-way power splitter, 50Ω 11667A Option 001
Attenuator, 10 dB 8491A Option 010
Attenuator, 30 dB 8491A Option 030
RF cable set, 50Ω type-N 11851B