Technical data
Chapter 2 2-67
Performance Tests
15. Source and Receiver Harmonics (Option 002 Only)
Procedure
Source Harmonics
1. Connect the equipment as shown in Figure 2-31.
Figure 2-31 Source Harmonics Test Setup
2. Press (or if the analyzer has Option 006) to set
the test port power to +20 dBm (or +18 dBm).
3. Press . To set the frequency range:
If you do not have Option 006, press .
If you have Option 006, press .
4. Press to set the IF bandwidth to 10 Hz.
5. Press .
6. After one sweep, press to normalize the
trace.
7. Press .
8. After one sweep, press to get a better viewing of the trace.
9. Press .
10.Write the marker 1 value (which appears on the analyzer display) on the performance
test record. This is the worst-case source 2nd harmonic.
11.To change the stop frequency:
If you do not have Option 006, press .
Preset Power 20 x1 18 x1
Start 16 M/µ
Stop 1.5 G/n
Stop 3 G/n
Avg
IF BW
10 x1
Meas
A
Display
DATA → MEMORY DATA/MEM
System
HARMONIC MEAS HARMONIC SECOND
Scale Ref
AUTO SCALE
Marker Search
SEARCH:MAX
Stop 1 G/n