Technical data
2-68 Chapter 2
Performance Tests
15. Source and Receiver Harmonics (Option 002 Only)
If you have Option 006, press .
12.Press .
13.After one sweep, press to normalize the
trace.
14.Press to get a better viewing of the trace.
15.Press .
16.After one sweep, press .
17.Press .
18.Write the marker 1 value on the performance test record.
Receiver Harmonics
19.Connect the equipment as shown in Figure 2-32.
Figure 2-32 Receiver Harmonics Test Setup
20.Press . To set the
frequency range:
If you do not have Option 006, press .
If you have Option 006, press .
21.Repeat steps 5 through 10 to measure the receiver worst case 2nd harmonic for the
receiver input B. Press in step 5.
Stop 2 G/n
System
HARMONIC MEAS HARMONIC OFF
Display
DATA → MEMORY DATA/MEM
Scale Ref
AUTO SCALE
System
HARMONIC MEAS HARMONIC THIRD
Scale Ref
AUTO SCALE
Marker Search
SEARCH:MAX
Power 6 x1 System
HARMONIC MEAS HARMONIC OFF
Stop 1.5 G/n
Stop 3 G/n
Meas
B