Technical data

2-70 Chapter 2
Performance Tests
16. Harmonic Measurement Accuracy (Option 002 Only)
16. Harmonic Measurement Accuracy
(Option 002 Only)
This test verifies the network analyzer’s accuracy when operating in the harmonic
measurement mode (Option 002). The analyzer’s reading is compared to that of the power
meter. The allowable difference is shown in the “Specifications” table.
Analyzer warmup time: 30 minutes
Specifications
Required Equipment
Frequency Range Harmonic Measurement
Accuracy
a
a. At 25 °C ± 5 °C, A, B, R, 10 dBm input.
16 MHz–3 GHz ± 1.5 dB
3 GHz– 6 GHz
b
b. Only for analyzers with Options 002 and 006.
± 3 dB
Description HP/Agilent Part or Model Number
External source 83620A or 83712A
Power meter 436A/437B/438A or E4418B/E4419B
Power sensor 8482A
2-way power splitter, 50 11667A Option 001
Attenuator, 10 dB 8491A Option 010
RF cable set, 50 type-N 11851B
Cable, BNC 24-inch 8120-1840
Adapter, 3.5-mm (f) to type-N (f) 1250-1745
Additional Equipment for Option 006 Analyzers
Power sensor 8481A