Technical data

Chapter 2 2-71
Performance Tests
16. Harmonic Measurement Accuracy (Option 002 Only)
Procedure
A Channel Harmonic Measurement Accuracy
1. Connect the equipment as shown in Figure 2-33.
Figure 2-33 Harmonic Measurement Accuracy Test Setup
2. Zero and calibrate the power meter. Set the power meter to measure dBm. Refer to the
power meter operating manual for more information on how to perform these tasks.
3. Preset the external source, and set the power level to 3.5 dBm and the CW frequency
to 32 MHz. Set the power meter for the appropriate calibration factor.
4. Write the power meter reading on the performance test record. It should read
approximately 10 dBm.
5. On the network analyzer, press to set
the analyzer internal source to 16 MHz.
6. Press .
7. Press . Write the marker 1 reading, which appears on the analyzer
display, in the “Input A Value” column on the performance test record.
8. On the external source, set the CW frequency to 48 MHz. Reset the power meter for the
appropriate calibration factor.
9. Record the power meter reading on the performance test record.
Preset Sweep Setup
CW FREQ
16 M/µ
System
HARMONIC MEAS SECOND
Marker Meas
A