Technical data

Chapter 2 2-77
Performance Tests
Performance Test Records
2. Source Power Range, Linearity, and Accuracy:
Power Range and Linearity (continued)
Source Power
Level (dBm)
Power Offset
(dB)
Measured
Value (dB)
Power
Linearity
(dB)
Spec (dB) Measurement
Uncertainty
CW Frequency = 3 GHz
5+15 ±0.25 ±0.02
3+13 ±0.25 ±0.02
1+11 ±0.25 ±0.02
+1 +9 ±0.25 ±0.02
+3 +7 ±0.25 ±0.02
+5 +5 ±0.25 ±0.02
+7 +3 ±0.25 ±0.0035
+9 +1 ±0.25 ±0.0022
+11 –1 ±0.25 ±0.0014
+13
a
–3 ±0.25 ±0.0009
+13
b
–3 ±0.5 ±0.0009
+15 –5 ±0.5 ±0.0006
+18 –8 ±0.5 ±0.17
+20
a
–10 ±0.5 ±0.17
a. Analyzers without Option 006 only.