Technical data

2-80 Chapter 2
Performance Tests
Performance Test Records
Agilent Technologies Company
Model 8753ES Option 011
Serial Number
Report Number
Option(s) Date
3. Receiver Minimum R Channel Level
Note: If your analyzer does not have Option 006, write “N/A” in all entries above 3 GHz.
CW Frequency Specification
(dBm)
Marker Value (dB) Measurement
Uncertainty
(dB)
300 kHz < 35 ±1.0
3.29 MHz < 35 ±1.0
3.31 MHz < 35 ±1.0
15.90 MHz < 35 ±1.0
16.10 MHz < 35 ±1.0
30.90 MHz < 35 ±1.0
31.10 MHz < 35 ±1.0
1.6069 GHz < 35 ±1.0
1.6071 GHz < 35 ±1.0
3.000 GHz < 35 ±1.0
4.000 GHz
a
a. Option 006 only.
< 30 ±2.0
5.000 GHz
a
< 30 ±2.0
6.000 GHz
a
< 30 ±2.0