User`s guide

1-108
Making Measurements
Using Test Sequencing
In-Depth Sequencing Information
Features That Operate Differently When Executed in a Sequence
The analyzer does not allow you to use the following keys in a sequence:
and keys
key, and
(backspace key)
Commands That Sequencing Completes Before the Next Sequence Command
Begins
The analyzer completes all operations related to the following commands before continuing
with another sequence command:
single sweep
number of groups
auto scale
marker search
marker function
data (
) memory
recall or save (internal or external)
copy list values and operating parameters
CHAN1, CHAN2, Wait 0*
*Wait 0 is the special sequencing function with a zero entered for the delay value.
Commands That Require a Clean Sweep
Many front panel commands disrupt the sweep in progress, for example, changing the
channel or measurement type. When the analyzer does execute a disruptive command in a
sequence, some instrument functions are inhibited until a complete sweep is taken. This
applies to the following functions:
autoscale
data memory
Forward Stepping in Edit Mode
In the sequence modify mode, you can step through the selected sequence list, where the
analyzer executes each step using the .
Titles
A title may contain non-printable or special ASCII characters if you download it from an
external controller. A non-printable character is represented on the display as π.
Preset
WAIT x