User`s guide

2-23
Making Mixer Measurements
High Dynamic Range Swept RF/IF Conversion Loss
Using the Mixer Measurement Diagram
While the analyzer is still set to the IF frequency range, press:
.
Note the RF frequency values on the diagram.
Press .
Perform a Power Meter Calibration Over the RF Range
1. Connect the equipment as shown in Figure 2-18.
Figure 2-18 Connections for Power Meter Calibration
2. Use the previous power meter settings.
3. Perform a one sweep power meter calibration over the RF frequency range at 5 dBm:
.
Perform the High Dynamic Range Measurement
1. Return the analyzer to the IF frequency range. Press
.
2. Make the connections shown in Figure 2-19.
System
INSTRUMENT MODE
FREQ OFFS MENU
LO MENU
FREQUENCY:CW
1500
M/µ
RETURN
RETURN
DOWN CONVERTER
RF > LO
Start 1.6 G/n Stop 2.5 G/n
Cal
PWRMTR CAL
ONE SWEEP
5 x1
TAKE CAL SWEEP
Start 100 M/µ Stop 1
G/n