User`s guide

2-47
Making Mixer Measurements
Isolation Example Measurements
5. Make the connections as shown in Figure 2-36.
Figure 2-36 Connections for a Response Calibration
6. Perform a response calibration by pressing
.
7. Make the connections as shown in Figure 2-37.
Figure 2-37 Connections for a Mixer RF Feedthrough Measurement
8. Connect the external LO source to the mixer’s LO port.
9. The measurement results show the mixer’s RF feedthrough.
NOTE You may see spurious responses on the analyzer trace due to interference
caused by LO to IF leakage in the mixer. This can be reduced with averaging
or by reducing the IF bandwidth.
Cal
CALIBRATE MENU
RESPONSE
THRU