User`s guide

4-48
Printing, Plotting, and Saving Measurement Results
Saving Measurement Results
Data Arrays
Press .
Data created the first time in this manner will be saved as filename “DATA00.d1.” The file
extension .d1 indicates that the data from the analyzer's channel 1 is error corrected data
only if the analyzer's error correction feature is enabled (in other words, you have
performed a calibration). Otherwise, the data is the same as data stored in the analyzer's
raw data arrays. Data stored in the data arrays does not have any formatting applied to it.
Format Arrays
Press
Data created the first time in this manner will be saved as filename “FILE00.f1”. The file
extension .f1 indicates the data is formatted per Figure 4-13 using the analyzer's channel
1. Depending on what features you've selected, data in the format arrays includes data in
the data arrays plus one or more of the following features:
Trace math (i.e data-memory)
Gating (Option 010)
Electrical delay
Conversion (for complex impedance (Z), admittance (Y), etc.)
Transform (Option 010)
Format (log, lin, phase, delay, SWR, excluding Smith and Polar)
Smoothing
In each of these examples, most users will select , under the
softkey menu. If is turned ON, an additional
file will be created with file extension .g0. This is a Hewlett-Packard Graphics Language
(HPGL) file.
Save/Recall
DEFINE DISK-SAVE
DATA ARRAY ON
Save/Recall
DEFINE DISK-SAVE
FORMAT ARY ON
SAVE USING ASCII
DEFINE DISK-SAVE
GRAPHICS on OFF