User`s guide
6-20
Calibrating for Increased Measurement Accuracy
Frequency Response and Isolation Error Corrections
If your type of calibration kit is not listed in the displayed menu, refer to "Modifying
Calibration Kits" on page 7-58.
6. To select a response and isolation correction and to start with the response portion of
the calibration, press:
7. Connect the short or open calibration standard to the port you selected for the test port
(PORT 1 for S
11
or PORT 2 for S
22
on ES analyzers).
NOTE Include any adapters that you will have in the device measurement. That is,
connect the standard device to the particular connector where you will
connect your device under test.
Figure 6-6 Standard Connections for a Response and Isolation Error
Correction for Reflection Measurements
8. To measure the standard, press:
or
If the calibration kit you selected has a choice between male and female calibration
standards, remember to select the sex that applies to the test port and not the standard.
The analyzer displays WAIT - MEASURING CAL STANDARD during the standard
measurement. The analyzer underlines the softkey that you selected after it finishes
the measurement, and computes the error coefficients.
9. Connect the load calibration standard to the test port.
CALIBRATE MENU
RESPONSE & ISOL’N
RESPONSE
SHORT
OPEN