User`s guide
6-43
Calibrating for Increased Measurement Accuracy
Calibrating for Noninsertable Devices
3. Connect adapter A3 (same sex and connector type as the DUT) to adapter A2 on port 2
as shown in Figure 6-15.
Figure 6-15 Two-Port Cal Set 1
4. Perform a full 2-port calibration between ports 1 and 2 using calibration standards
appropriate for the connector type at port 1 (the connector type for adapter A1). Save
the calibration by selecting . Name the file "PORT1."
5. Connect adapter A3 to adapter A1 on port 1 as shown in Figure 6-16.
Figure 6-16 Two-Port Cal Set 2
6. Perform a full 2-port calibration between ports 1 and 2 using calibration standards
appropriate for the connector type at port 2 (the connector type for adapter A2). Save
the calibration by selecting . Name the file "PORT2."
NOTE In the following steps, calibration data is recalled, not instrument states.
Save/Recall
SAVE STATE
Save/Recall
SAVE STATE